產品概述:
化(hua)學(xue)耗氧(yang)量(又稱COD)是指在一(yi)定(ding)(ding)條件下,用強氧(yang)化(hua)劑將水樣中(zhong)有機物(wu)和無(wu)機還(huan)原(yuan)性物(wu)質氧(yang)化(hua)時所消耗氧(yang)化(hua)劑相對應氧(yang)的質量濃度(du)。COD也(ye)是反映水體受有機物(wu)和無(wu)機還(huan)原(yuan)性物(wu)質污染(ran)程度(du)的重要(yao)指標(biao)(biao)。該(gai)分析儀器產品基于港能公司水質分析測(ce)(ce)試專利技術,依據(ju)國家標(biao)(biao)準測(ce)(ce)試方法《水質-化(hua)學(xue)耗氧(yang)量測(ce)(ce)定(ding)(ding)-重鉻酸鉀(jia)(GB11914-89)》研制(zhi)設計,由流(liu)體動力單(dan)(dan)元(yuan)、可(ke)視(shi)光電定(ding)(ding)量單(dan)(dan)元(yuan)、流(liu)路切換單(dan)(dan)元(yuan)、消解測(ce)(ce)量單(dan)(dan)元(yuan)及嵌入(ru)式(shi)控制(zhi)器單(dan)(dan)元(yuan)組成,可(ke)快速、準確(que)、連續、可(ke)靠完成水體中(zhong)COD的測(ce)(ce)定(ding)(ding)。
該分析儀器(qi)能夠根(gen)據現(xian)(xian)場(chang)設定長期無人值守自動連續工作(zuo),廣泛適用(yong)于工業(ye)(ye)污(wu)(wu)染源(yuan)排放廢(fei)水(shui)(shui)、工業(ye)(ye)過程(cheng)工藝廢(fei)水(shui)(shui)、工業(ye)(ye)污(wu)(wu)水(shui)(shui)處(chu)理廠污(wu)(wu)水(shui)(shui)、市政污(wu)(wu)水(shui)(shui)處(chu)理廠污(wu)(wu)水(shui)(shui)等場(chang)合。根(gen)據現(xian)(xian)場(chang)測試(shi)工況的復雜程(cheng)度,可選(xuan)配相應的預(yu)處(chu)理系統(tong),以(yi)保證測試(shi)過程(cheng)可靠、測試(shi)結果(guo)準確,充分滿足(zu)不同場(chang)合現(xian)(xian)場(chang)需求(qiu)。
基本原理
水樣、重鉻酸鉀消解溶液、硫酸汞(硫酸汞可以消除水樣中氯離子的干擾,因氯離子能與汞離子形成非常穩定的氯化汞)、硫酸銀溶液(硫酸銀作為催化劑加入可以更有效地氧化直鏈脂肪化合物)混合液加熱到147℃,重鉻酸離子氧化溶液中的有機物后顏色會發生變化,分析儀檢測此顏色的變化,并把這種變化換算成COD值輸出出來。消耗的重鉻酸離子量相當于可氧化的有機物量。
技術規格
方法依據:國家標準GB11914-89《水質-化學耗氧量測定-重鉻酸鉀》
測量范圍:4-50000 mg/L COD
準確度: ≥100mg/L時,不超過±10%;<100mg/L時,不超過±8mg/L
重復性: ≥100mg/L時,不超過±10%;<100mg/L時,不超過±6mg/L
測量周期:小測量周期為40分鐘,據實際水樣,可在5~120min任意修改消解時間
采樣周期:時間間隔(40~9999min任意可調)和整點測量模式
校準周期:1~99天任意間隔任意時刻可調
維護周期:一般每月一次,每次約30 min
試劑消耗:小于0.35元/樣品
輸出:RS-232
環境要求:溫度可調的室內,建議溫度+5~28℃;濕度≤90%(不結露)
電源:AC230±10%V,50±10%Hz,5A
尺寸:高1500×寬550×深450mm
其(qi)他:異(yi)常報警和斷電(dian)不會丟(diu)失數(shu)據(ju);觸摸(mo)屏顯示(shi)及指令輸入(ru)
|